HTS is an electronic module assembled inside Perfomer™ (click here). It collects the SVO2 and the Hct% data transferred by BLOP4 or BLOP4-CPB and send them to Performer Central Processing Unit through a RS-232 serial interface.
This board acquires 8 external indipendent temperatures coming from 8 thermistor probes measuring the temperature of a fluid flowing inside a disposable plastic cuvette.
Legal Manufacturer/Customer: RanD S.r.L.
Application: visit RanD web site
Datamed srl role in the project: Hardware and Firmware designer and board manufacturer
Datamed S.r.L. is the designer and the manufacturer of the electronic and firmware of Data Master (see here).
Data Master performs oxymetry (Sat%), heamatocrit (Hct%), blood temperatures and arterial PO2 measurements in a single unit.
Arterial PO2 range: 0-600 mmHg
PO2 resolution: 1 mmHg
PO2 accuracy: 3% PO2 0-200 mmHg, 5% PO2 > 200 mmHg
Temperature range: 10-45 °C
Temperature resolution: 0.1 °C
Temperature accuracy: 1%
Venous Sat% range: 40-100%
Sat% resolution: 0.1%
Sat% accuracy: 2% between 60-100%; 4% between 40-60%
Hct% range: 15-50%
Hct% resolution: 1%
Hct% accuracy: 3% between 15-30%; 4% between 30-50%
Temperature range: 10-45 °C
Temperature resolution: 0.1 °C
Temperature accuracy: 1%
Legal Manufacturer/Customer: SORIN GROUP
Application: Cardiac Surgery